Welcome to Liquid International — factory-direct lab & training equipment from China

LI-QUID LIQUID INTERNATIONAL 利奎德国际 · Shanghai
Inquiry Basket0

元素分析仪

Elemental Analyzers

Model No. ELA-0001 · Elemental Analysis Instruments

Combustion CHNS/O Organic Elemental Analyzer

燃烧式 CHNS/O 有机元素分析仪

Synthetic chemists, polymer researchers and MOF groups confirm composition by combustion CHNS, with oxygen as an option. This analyzer burns 1–10 mg at 950–1150 °C, detects by TCD and reports each element from 0.01% to 100%, within 0.3% absolute on pure standards. Acetanilide repeatability is ≤0.2% RSD with 5–10 min runs.

Elements measured
C, H, N, S (O optional)
Sample mass
1–10 mg
Detection range
0.01–100% per element
Accuracy on pure standards
Within 0.3% absolute
Repeatability (RSD)
≤0.2% for acetanilide
Inquiry Now

Packing & Shipping

Place of originChina (dispatched from Shanghai)
BrandLiquid International / OEM on request
Export packingFumigation-free plywood case or reinforced carton
Loading portShanghai / Ningbo (others on request)
Payment termsT/T, L/C at sight (confirmed on quotation)
Trade termsEXW, FOB, CFR, CIF, DAP

Typical export terms — final terms are confirmed on the proforma invoice.

Product Description 产品描述

Synthetic chemists, polymer researchers and MOF groups confirm composition by combustion CHNS, with oxygen as an option. This analyzer burns 1–10 mg at 950–1150 °C, detects by TCD and reports each element from 0.01% to 100%, within 0.3% absolute on pure standards. Acetanilide repeatability is ≤0.2% RSD with 5–10 min runs.

Highlights

  • Accuracy within 0.3% absolute for publication-grade results
  • Optional oxygen determination
  • Small samples suit research compounds
  • Thermal conductivity detection, simple to maintain

Applications

  • Organic compound purity confirmation
  • Pharmaceutical intermediate testing
  • Polymer composition
  • Catalyst and MOF characterization

Specifications 技术参数

ParameterValue
Elements measured 测定元素 C, H, N, S (O optional)
Sample mass 样品量 1–10 mg
Detection range 检测范围 0.01–100% per element
Accuracy on pure standards 纯标样准确度 Within 0.3% absolute
Repeatability (RSD) 重复性(RSD) ≤0.2% for acetanilide
Combustion temperature 燃烧温度 950–1150 °C
Run time 运行时间 5–10 min
Detector 检测器 Thermal conductivity detector
Carrier gas 载气 Helium, 99.999%
Autosampler 自动进样器 50–120 positions
Model No.ELA-0001
Place of originChina (dispatched from Shanghai)

Applicable standards 适用标准

ASTM E1019 ASTM D5373 ASTM D5453 ASTM D4629 ISO 10304-1 ASTM D7359 ISO/IEC 17025 IEC 61010-1

Chinese national standards (GB/T, JJG) and international standards referenced for this product type. Test reports and certificates are supplied with the quotation on request.

Packing & Shipping 包装与运输

Export packing

Fumigation-free plywood case or reinforced carton

Every unit is power-tested before packing; enclosed in shock foam and moisture barrier; wooden cases comply with ISPM 15.

Loading port

Shanghai / Ningbo (others on request)

Full container load (FCL), less-than-container load (LCL) or air freight; complete export documentation provided.

Inquiry Now 索取报价

Complete the form below — an export engineer will send technical configurations and a formal proforma invoice within 24 hours.

Products in this inquiry: Combustion CHNS/O Organic Elemental Analyzer (ELA-0001)

We reply within 24 hours (Beijing Time, UTC+8). Your details are used only to answer this inquiry.

Related products

5 models

Send your equipment list — receive a formal quotation within 24 hours

Single instruments, complete laboratories or government tenders — our export engineers prepare the configuration, proforma invoice and shipping documents.